11. Click Activate. At the Install License screen, click OK.
CAUTION Do not disconnect or power off the device during
license installation.
12. Once you see the Add new features screen confirming the license installation, click
Close to close the Service Tool window.
13. For the license upgrade to take effect, press and hold the power button for about 8
seconds to reboot the device.
Upgrade your existing CP 150 device wireless radio
software version through the Welch Allyn Service Tool
Note A USB cable is required to perform the wireless radio software upgrade.
1. Download the Welch Allyn Service Tool and Partner Connect and install on your PC.
2. Plug your CP 150 device into a USB port on your PC running the Welch Allyn Service
Tool and turn on the device.
3. If the Welch Allyn Service Tool is not still open, go to the Windows Start menu, and
then select All Programs > Welch Allyn > Welch Allyn Service Tool.
4. Log in to the Welch Allyn Service Tool.
5. Highlight the CP150 to choose it from the Device list.
6. Click Select.
44 Service CP 150 12-lead resting electrocardiograph