9. Once you see the Upgrading Welch Allyn 802.11 a/b/g/n Radio... screen confirming
the completed installation, click Close.
10. For the wireless radio upgrade to take effect, press and hold the power button for
about 8 seconds to reboot the device.
Configure your network policy certificate through the
Welch Allyn Service Tool
Note A USB cable is required to perform the network policy configuration.
1. Download the Welch Allyn Service Tool and Partner Connect and install on your PC.
2. Plug your CP 150 device into a USB port on your PC running the Welch Allyn Service
Tool and turn on the device.
3. If the Welch Allyn Service Tool is not still open, go to the Windows Start menu, and
then select All Programs > Welch Allyn > Welch Allyn Service Tool.
4. Log in to the Welch Allyn Service Tool.
5. Highlight the CP150 to choose it from the Device list.
6. Click Select.
46 Service CP 150 12-lead resting electrocardiograph