Restore factory defaults and export or import of the device
configurations
Restore factory defaults
Button Feature
All Settings
Restores every setting to the factory default setting.
Printer page count
Restores the printer page to 0. Use this feature after the printer assembly is
replaced.
Calibration gain
Resets the spirometry calibration gain to 1.0
Radio settings
Restores every radio setting to the factory wireless radio default setting.
Device configuration
Configuration files can be exported and imported for backup purposes or to configure
multiple CP150 devices.
Button Feature
Save to USB
Saves the device configuration to USB flash drive.
Configure from USB
Loads the device configuration from USB flash drive.
Print all settings
Prints a list of all current CP 150 device settings to the device printer.
Update
Updates the device software from a USB flash drive.
To begin the export or import process, insert a USB flash drive into one of the USB
ports on the back of the CP150 device.
1.
From the Home screen, touch the Settings tab. The ECG tab and the vertical ECG
configuration tab appear.
2. Touch the Advanced tab.
3. Enter 6345 as the access code and touch OK.
The General tab appears at the bottom of the screen and the Regional tab appears at
the top of the screen.
4. Touch the Service tab.
Touch Save to USB to save the device configuration to USB.
Touch Configure from USB to copy the configuration from USB.
Touch Update to install the latest device software.