Symptom Possible cause Suggested action
when high-speed printing has not
taken place.
The read-back file is corrupted. During the Service USB Host test, the content
read back from file does not match the content
previously written.
Ensure that the thumb drive and
port are working properly. If
using another thumb drive results
in the same error, contact
Technical Support.
Internal printer is out of paper.
Reload.
The cue marker on the paper sensor is causing
the printer to incorrectly identify that paper tray
is empty.
Check Que Sensor board and
harness.
Post test error messages
Post test error messages:
Symptom Possible cause Suggested action
System Fault
code 01
Battery capacity is at critical capacity and auxiliary power is
not present
Ensure that AC power is plugged in.
Check for faulty battery and/or 60W power
supply.
Check all harness to power supply.
System Fault
code 02
Core image verification fails Reload software through the Welch Allyn
Service Tool.
System Fault
code 03
Core image is missing Reload software through the Welch Allyn
Service Tool.
System Fault
code 04
Software update failure Reload software through the Welch Allyn
Service Tool.
System Fault
code 05
RAM Test failure Check for faulty SMCE board.
66 Troubleshooting CP 150 12-lead resting electrocardiograph