62 Chapter 5 Performing ECG Tests Welch Allyn CP 200 Electrocardiograph
Recording a Stat Auto ECG
A stat Auto ECG is an immediate printout in Auto Report 1 format.
Stat mode bypasses patient data entry, as shown in
Figure 36 on page 56. A temporary
ID number is assigned to the patient to identify stat tests. After printing, you may enter
the patient data by editing the test.
In stat mode, Auto Send and Auto Save are always disabled, even if they are enabled in
the ECG settings. If you want to send or save a stat Auto ECG, you may do so manually
after you print.
To Record a Stat Auto ECG
1. Press and hold
the Auto ECG key .
The electrocardiograph begins acquiring ECG data. After it has acquired 10 seconds of
quality data, it prints a report.
2. Go to
Step 7 on page 58.
Continue the procedure as if you have just pressed the Print ECG softkey.