EasyManua.ls Logo

Welch Allyn CP 200 - Testing the Equipment

Welch Allyn CP 200
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Directions for Use Chapter 7 Maintaining the Electrocardiograph 81
Testing the Equipment
Whenever the electrocardiograph is serviced or problems are suspected, Welch Allyn
recommends the following test procedures:
1. Verify continued electrical safety of the device, using IEC 60601-1 or ANSI/AAMI ES1
methods and limits. Test for the following:
Patient leakage current
Chassis leakage current
Earth leakage current
Dielectric strength (AC and patient circuits)
2. Verify that the electrocardiograph is working properly, using an ECG simulator to
acquire and print a standard 12-lead ECG of known amplitude.
Printing should be dark and even across the page.
There should be no evidence of print-head dot failure (no printing breaks forming
horizontal streaks).
Paper should move smoothly and consistently during printing.
Waveforms should appear normal, with proper amplitude, and without distortion
or excessive noise.
Paper should stop with perforations near the tear bar, indicating proper cue-
sensor operation. For tear-bar location, see Figure 17 on page 23.
WARNING Only qualified service personnel should perform leakage current
tests.

Table of Contents

Other manuals for Welch Allyn CP 200

Related product manuals