Item Specification
2x6 + 1R @ 50 mm/s, No print
Report formats, external
printer*
Auto ECGs
Average cycles
3x4 @ 25 mm/s, 3x4 @ 50 mm/s, 3x4 + 1R @ 25 mm/s, 3x4 + 3R @ 25 mm/s,
3x4 - 5.0s @ 25 mm/s, 3x4 - 5.0s @ 50 mm/s, 6x2 - 5.0s @ 25 mm/s,
12x1 - 5.0s @ 25 mm/s, No print
3x4 + 3R @ 25 mm/s, 3x4 + 3R @ 50 mm/s, 2x6 + 1R @ 25 mm/s,
2x6 + 1R @ 50 mm/s, 3x4 + 2R @ 25 mm/s, 3x4 + 2R @ 50 mm/s,
6x2 + 2R @ 50 mm/s, No print
ECG storage (in test
directory)
At least 25 ECG tests
Patient storage (in
worklist, CP 50 Plus only)
Up to 50 patients
Frequency range 0.3 to 150 Hz
Digital sampling rate > 1,000 samples/second/channel
Pacemaker detection ANSI/AAMI EC11
Power requirement Universal AC power supply 110-240 V~, 50/60 Hz ~, 1.5 A maximum
AC fuses Time-lag type, 2.0-amp 250-V rating, Littlefuse 0218002P or equivalent
Rechargeable battery 10.8V, 1.9Ah minimum, 3-cells Li-ion “smart” battery. A fully charged battery prints up to 25
ECGs with one extra copy per test. Recharge time to 90 percent capacity is 6 hours.
Filters
High-performance
baseline
Muscle tremor
AC interference
0.5 Hz
35 Hz
50 Hz or 60 Hz
Safety, EMC, and
regulatory compliance
ANSI/AAMI EC11**
CAN/CSA C22.2 No. 601.1
CAN/CSA C22.2 No. 601.1.1
CAN/CSA C22.2 No. 601.1.2
CAN/CSA C22.2 No. 601.1.4
CAN/CSA C22.2 No. 601.2.25
UL60601-1
IEC/EN 60601-1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-1-6
IEC/EN 60601-2-25 ***
IEC/EN 60601-2-51 (2x6 lead arrangement)
Standard connectivity Mini USB client and USB host, Ethernet (optional - available with CP 50 Plus only)
44 Specifications CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph