Guidance and manufacturer’s declarations
EMC compliance
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment. This device complies with IEC EN 60601-1-2:2007.
All medical electrical equipment must be installed and put into service in accordance
with the EMC information provided in this document and the Welch Allyn website:
http://www.welchallyn.com/apps/products/product_category.jsp?catcode=CARDIO.
Portable and mobile RF communications equipment can affect the behavior of
medical electrical equipment.
The electrocardiograph complies with all applicable and required standards for
electromagnetic interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is not safe to operate the electrocardiograph in the presence of high-frequency
surgical equipment.
However, it is good practice to avoid using the electrocardiograph in extremely close
proximity to other equipment.
Emissions and immunity information
Electromagnetic emissions
The Welch AllynCP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph is intended for use in the electromagnetic
environment specified below. The customer or user of the Welch AllynCP 50™ and CP 50 Plus™ 12-lead resting
electrocardiograph should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The Welch AllynCP 50™ and CP 50 Plus™ 12-lead resting
electrocardiograph uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Welch AllynCP 50™ and CP 50 Plus™ 12-lead resting
electrocardiograph is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low
47