4
KHz
ANSI Standard RTL for
0.
dB
HL
=
9.5
dB
SPL
Reference HL for Calibration
(+)80.
dB
HL
=
(+)80.0
dB
SPL
(+) or (-) Microphone Correction
NA.
(+)
1.0
dB
SPL
80.
dB
HL
=
90.5
dB
SPL
The following table contains the ANSI Standard Reference Threshold Levels
RTL's
at
0 HL for each frequency when using TDH-39P earphones.
Frequency
(Hz)
125
250
500 750 1000 1500
2000 3000
4000
6000 8000
ANSI
Standardd
RTL
45.0 25.5
11.5
7.5
7.0 6.5
9.0
10.0
0.5
15.5 13.0
Measured
in
dB
SPL
Standard Reference Threshold Levels re:
20pPa
for
Telephonics
TDH-39P earphones
as measured on the National Bureau of Standards 9-A coupler. Reference ANSI S3.6
-
1989, ISO 389
-
1975 Standards.
Grason-Stadler has made an effort to minimize calibration time by defaulting the
Reference HL Levels for Calibration to maintain an expected SPL Calibration Level of
between 85 and 95
dB
SPL. The Default HL Levels are as follows:
Frequency
(Hz)
125 250 500 750 1000 1500
2000 3000
4000
6000
8000
Default HL
40
60 80 80 80 80
80 80 80 80 60
-
Because the 38 automatically adds the ANSI RTL and reference HL values on the
display (when the Tone Bar, is active) the technician needs only to add or subtract
the appropriate microphone correction to the displayed value when calibrating.
3)
4)
5)
6)
7)
8)
Connect the right ear phone to the Sound Level Meter Artificial Ear.
Press (TONE BAR\ the display will update from the selected hearing level to
the ANSI Standard RTL value for calibration and the
m+
indicator will
appear).
Adjust the HL knob until the SPL Level measured on the SLM for the selected
frequency and transducer equals the value indicated on the 38 display (+ or
-
Microphone Correction).
Press
theIF
Key to store the data in memory.
Repeat for all frequencies, right phone, left phone, ipsi and contra by
selecting the appropriate key on the 38 front panel.
NOTE: Both left phone and right phone must be calibrated.
When finished press the
vl
Key to return to the Main Calibration Mode
Menu.