Calibration
Calibration
Procedure
To
calibrate
the
unit
proceed
as
follows:
Connect
the
calibration
pump
to
the
sensor.
Ensure
that
there
are
no
air
leaks
Press
the
CAL
key
-
the
menu
is
displayed
as
shown
on
the
previous
page
The
cursor
is
positioned
at
temperature.
Enter
the
ambient
temperature
Press
ENTER
when
the
correct
temperature
entered
Wait
1
second.
Press
the
START
key
Pump
4
to
6
litres
of
air
through
the
sensor
Note:
Make
sure
that
the
flow
sensor
is
kept
still
during
the
pumping
operation.
While
pumping,
the
unit
records
the
volume
being
pumped
through
the
flow
sensor
and
indicates
it
on
the
display.
Press
the
STOP
key
when
finished
pumping,
Enter
the
pumped
air
volume
at
syringe
volume
prompt
(depending
on
the
size
of
the
calibration
pump
and
times
of
pumping;
i.e.
a
2
litre
pump
pumped
3
times
*
6
litres).
Press
ENTER.
The
message
"Calibration
complete
"
appears
on
the
display,
press
the
PRINT
key
to
obtain
a
printout
of
die
calibration
report
with
the
following
information:
Page
22
SP-1
User's
Guide