36 Program Mode Welch Allyn TM286 Auto Tymp
Figure 15. Display format for TYMP/REFLEX Test
(Reflex test results given as Yes or No)
When the reflex test cannot be performed, due to a leak or early extraction of the probe, a
“NT” will appear next to the frequency.
To select a different setting for reflex format:
1. While in the Program mode, move the cursor to the desired setting.
2. While the square cursor is positioned in front of the desired setting, press the button.
The word SAVED appears in the lower right corner of the screen. The previous setting is
deselected. An asterisk (*) is displayed beside the new default setting.
REFLEX
This feature determines the stimuli and signal routing of the acoustic reflexes as default
settings. To select the frequencies, move the cursor next to the selection and press the
button to save. Saved will appear on the bottom right corner of the LCD. An * will
appear next to the selected stimulus routing and frequency. The system will allow 4
stimulus choices of any combination (i.e., ipsi or contra) for display and printing.
AUTO HL SETUP
Programming the Auto HL Procedure
Navigate the cursor to the Auto HL Setup line located on the Program Mode Screen 1
and press the button. The following submenu will appear:
Ipsi 500
Ipsi 1000
Ipsi 2000
Ipsi 4000
M +
Test Frequencies (Hz) . . .
Intensity Range (dB HL) . . .
Start test ear . . .