58 Computer Interface Welch Allyn TM286 Auto Tymp
During Printing
NO DATA AVAILABLE
This message appears if the button is pressed and no results are stored.
NOT AVAILABLE
This message appears if the button is pressed and the computer is not properly
connected.
Data Transfer Program Mode
The Data Transfer Program mode is used to modify the TM286 USB interface
configuration parameters to match the computer's USB Port settings.
Enter the Program mode by selecting the Program (PROG) button. Move the cursor to
Data Xfer Config and press the button to enter the submenu.
The following screen appears the first time the Data Transfer Program mode is entered
showing the factory default settings.
These selections fall into three groups:
•Baud rate
Parity and data bits
Flow control
The default setting for each group has an asterisk (*) before it so that it is easy to scan the
settings for each group.
Selecting default settings for any of the groups is achieved in the same manner as the
Program mode. Use the or buttons to move the solid square cursor down or up
to the setting that you wish to select. Press the button. The word SAVED will appear in
the lower right corner of the screen and the asterisk (*) will appear in front of the new
setting.
To exit the Data Transfer Program submenu, move the cursor to the press the
button. This will return to the Program mode which can be exited by selecting Program
(PROG).
PROGRAM MODE - DATA TRANSFER
* 115.2 kBAUD * NO PARITY + 8-BIT
57.6 kBAUD ODD PARITY + 7-BIT
38.4 kBAUD EVN PARITY + 7-BIT
19.2 kBAUD SPC PARITY + 7-BIT
9600 BAUD * XON/XOFF DISABLED
4800 BAUD XON/XOFF ENABLED