EasyManua.ls Logo

Welch Allyn VSM 6000 series - Guidance and Manufacturers Declaration; EMC Compliance; Emissions and Immunity Information

Welch Allyn VSM 6000 series
143 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Guidance and manufacturer's declaration
EMC compliance
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical
electrical equipment. This device complies with IEC EN 60601-1-2:2007.
All medical electrical equipment must be installed and put into service in accordance with the
EMC information provided in this document and the VSM 6000 Series Directions For Use.
Portable and mobile RF communications equipment can affect the behavior of medical
electrical equipment.
The VSM 6000 monitor complies with all applicable and required standards for electromagnetic
interference.
It does not normally affect nearby equipment and devices.
It is not normally affected by nearby equipment and devices.
It is not safe to operate the monitor in the presence of high-frequency surgical equipment.
However, it is good practice to avoid using the monitor in extremely close proximity to other
equipment.
Emissions and immunity information
Electromagnetic emissions
The monitor is intended for use in the electromagnetic environment specified below. The customer or user of
the monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The monitor is suitable for use in all establishments other than
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning is
heeded:
Harmonic emissions
IEC 61000-3-2
Class A
105