Message Possible cause Suggested action
Replace the SpO2 sensor. The SpO2 sensor is faulty Replace the SpO2 sensor.
No SpO2 sensor is connected Connect an SpO2 sensor.
Temperature messages
Message Possible cause Suggested action
Connect temperature probe. No probe is connected Connect a temperature probe
and retry.
The probe is faulty Replace the temperature
probe.
The temperature module returned a connect
probe message
Connect a temperature probe
and retry. If a probe is already
connected, replace the probe.
Insert correct color-coded
probe well.
The probe well is missing Insert a temperature probe
well.
Replace temperature probe. The probe is faulty Replace the temperature
probe.
Temperature not functional.
Call for service.
A module or messaging error occurred Call for service.
The ear thermometer dock is disconnected
from the monitor
Connect the dock to the
monitor.
Temperature time limit
exceeded.
The 10-minute timeout for temperature
measurement has occurred
Remove the probe from the
measurement site.
Tissue contact lost The probe has lost contact with the patient's
tissue.
Reposition the probe to
restore proper contact with
the patient's tissue.
Retry temperature
measurement.
Note
This message
often
accompanies
other
temperature
messages.
A probe heater or data error occurred Retry the temperature
measurement. If the problem
persists, replace the probe.
User settings require adjustment Adjust the user settings and
retry.
Patient data management messages
Message Possible cause Suggested action
Maximum number of patient
records saved. Oldest record
overwritten.
The maximum number of patient records in
the monitor's memory has been exceeded
On the Review tab, delete old
records to prevent the alarm
Directions for use Troubleshooting 123