Alarm messages Priority
Alarm limit exceeded. Temperature HIGH. Low
Alarm limit exceeded. Temperature LOW. Low
Technical alarms
Alarm messages Priority
Low battery 5 minutes or less remaining. High
Searching for pulse signal. High
Communications module did not power on properly. Power down the device. High
Network not found; check network cable connections. Low
Powering down. Call for service. Low
Battery is absent or faulty. Call for service. Low
NIBP air leak; check cuff and tubing connections. Low
NIBP not functional. Call for service. Low
Unable to determine NIBP; check connections and tubing for kinks. Low
Incorrect NIBP cuff size; check patient type. Low
Inflation too quick; check NIBP cuff and tubing connections. Low
Unable to determine NIBP; check inflation settings. Low
SpO2 not functional. Low
Attach SpO2 sensor to monitor. Low
Replace the SpO2 sensor. Low
Set date and time. Low
Maximum number of patient records saved. Oldest record overwritten. Low
Unable to access patient information. Low
Connect temperature probe. Low
Insert correct color coded probe well. Low
Replace temperature probe. Low
Temperature not functional. Call for service. Low
Directions for use Alarms 55