Note Disinfect according to your facility's protocols and standards or local regulations.
CaviWipes or Sani-Cloth Plus
1. Using CaviWipes or Sani-Cloth Plus, wipe the surface of the monitor to remove all debris.
2. Allow the monitor surface to dry for a minimum of 10 minutes before using the monitor.
70 percent isopropyl alcohol
Wipe the monitor with a clean cloth slightly dampened with 70 percent isopropyl alcohol.
10 percent chlorine bleach solution
1. Wipe the monitor with a clean cloth slightly dampened with a 10 percent bleach and water
solution. Follow the cleaning agent manufacturer's guidelines.
2. Rinse with a clean cloth slightly dampened with water that meets EP and USP quality
standards.
3. Allow the monitor surface to dry for a minimum of 10 minutes before using the monitor.
Clean the accessories
1. Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild detergent
solution.
2. Wipe the temperature probe with a cloth dampened with alcohol, warm water, or an
appropriately diluted, nonstaining disinfectant solution.
3. Clean the SpO2 sensor with a cloth dampened with 70 percent isopropyl alcohol or 10 percent
chlorine bleach solution.
4. Clean the ear thermometer according to the manufacturer's directions for use.
5. Clean the ear thermometer dock with the same cleaning agents used on the monitor.
Caution Never immerse any monitor accessories, such as NIBP hoses or cuffs,
temperature probes, or SpO2 sensors.
Clean the stand
The same cleaning agents used on the monitor can be used on the stand and accessory cable
management system. Clean on a routine basis according to your facility's protocols and
standards, or local regulations.
Note Disinfect according to your facility's protocols and standards, or
local regulations.
Directions for use Maintenance and service 91