Do you have a question about the wilks InfraCal CVH and is the answer not in the manual?
Method | Infrared absorption |
---|---|
Resolution | 1 ppm |
Battery Life | 8 hours |
Communication Interface | USB |
Wavelength | 3.4 µm |
Sample Type | Water |
Accuracy | +/- 2% |
Response Time | <10 seconds |
Detector | Pyroelectric |
Power | Rechargeable battery |
Measurement Principle | Infrared |
Measurement Time | <10 seconds |
Display | Digital |
Operating Temperature | 0 to 45°C |
Storage Temperature | -20 to 60°C |
Properties of the calibration oil used for the analyzer, including heavy hydrocarbon content and absence of additives.
Steps to prepare the first calibration standard at 100 ppm concentration.
Steps to prepare the second calibration standard at 50 ppm concentration.
Steps to prepare the third calibration standard at 25 ppm concentration.
Detailed steps for zeroing the InfraCal using clean solvent.
Instructions for running standards multiple times and averaging results for calibration.
Process for entering standard absorbance and calibration values into the EDIT program.
Confirmation that the calibration is stored and active in the InfraCal.