wilks InfraCal CVH User Manual

wilks InfraCal CVH
35 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
InfraCal
TOG/TPH Analyzer,
Model CVH
Calibration
www.WilksIR.com
Question and Answer IconNeed help?

Do you have a question about the wilks InfraCal CVH and is the answer not in the manual?

wilks InfraCal CVH Specifications

General IconGeneral
MethodInfrared absorption
Resolution1 ppm
Battery Life8 hours
Communication InterfaceUSB
Wavelength3.4 µm
Sample TypeWater
Accuracy+/- 2%
Response Time<10 seconds
DetectorPyroelectric
PowerRechargeable battery
Measurement PrincipleInfrared
Measurement Time<10 seconds
DisplayDigital
Operating Temperature0 to 45°C
Storage Temperature-20 to 60°C

Summary

Calibration Oil

Calibration Oil Characteristics

Properties of the calibration oil used for the analyzer, including heavy hydrocarbon content and absence of additives.

Mixing Standards – Volume/volume

Creating 100 ppm Standard

Steps to prepare the first calibration standard at 100 ppm concentration.

Creating 50 ppm Standard

Steps to prepare the second calibration standard at 50 ppm concentration.

Creating 25 ppm Standard

Steps to prepare the third calibration standard at 25 ppm concentration.

Instrument Calibration Procedure

Zeroing the InfraCal Analyzer

Detailed steps for zeroing the InfraCal using clean solvent.

Running and Averaging Standards

Instructions for running standards multiple times and averaging results for calibration.

Entering Calibration Data

Process for entering standard absorbance and calibration values into the EDIT program.

Calibration Completion Status

Confirmation that the calibration is stored and active in the InfraCal.