What to do if my Wyatt Technology Laboratory Equipment shows a Low Light Alarm?
- CChristina DavisAug 3, 2025
If your Wyatt Technology Laboratory Equipment Optilab rEX shows a Low Light Alarm, several factors could be at play. It might be due to a dramatic change in the sample refractive index, like injecting a concentrated solution directly. To fix this, ensure you're using degassed solvents to prevent air bubbles. Also, check that the LED/fiber light source is turned on. If the light source is on but the photodiode array voltage is low, a bubble in the flow cell may be the cause.