Remote Operation
GPIB Operation
Release 3.0 121
Configuration
Configure the power supply’s GPIB address and power-on service request setting.
The defaults are GPIB address 2 and power-on service request off.
Front panel
First set the GPIB parameters:
Next, select GPIB as the remote control interface.
Step # Do This You Will See
1 ERROR MSGS
2 5 times REMOTE CONFIG
3 RS-232 Cfg
2 times GPIB Cfg
4 GPIB Addr ##
5 Select an address from 1-30 GPIB Addr 2
6
PON SRQ? Y
7 or
Select whether power-on service
request is to be sent.
Options are
Y (Yes) or
N (No)
8
Settings are saved.
Step # Do This You Will See
1 ERROR MSGS
2 4 times REMOTE SELECT
3 RS-232
4 or
Select GPIB.
GPIB
5
GPIB setting is saved.
MENU
ENTER
ENTER
ENTER
ENTER
MENU
ENTER
ENTER