Remote Operation
Remote Analog Programming of Output Voltage and Current Limit
Revision A 69
Table 4.1 Power Supply Settings for Different Programming Sources
Output Voltage
Programming
Source
Output Current Limit Programming Source
0-5 Vdc 0-10 Vdc 0-5 k Resistor 0-10 k Resistor
None (Front
Panel Control)
0-5 Vdc
SW1: use
default
settings.
J2: jumper
3 to 4, 3 to 5.
SW1: set
4 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
2 closed.
J2 Jumper
3 to 4, 3 to 5.
SW1: set
2 and 4 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: use
default settings.
J2: jumper
3 to 4.
0-10 Vdc
SW1: set
3 closed:
J2: jumper
3 to 4, 3 to 5.
SW1: set
3 and 4 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
2 and 3 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
2, 3, and 4
closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
3 closed.
J2: jumper
3 to 4.
0-5 k Resistor
SW1: set
1 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1 and 4 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1 and 2 closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1, 2, and 4
closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1 closed.
J2: jumper
3 to 4.
0-10 k Resistor
SW1: set
1 and 3
closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1, 3, and 4
closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1, 2, and 3
closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1, 2, 3, and 4
closed.
J2: jumper
3 to 4, 3 to 5.
SW1: set
1 and 3 closed.
J2: jumper
3 to 4.
None (Front
Panel Control)
SW1: use
default
settings.
J2: jumper
3 to 5.
SW1: set
4 closed.
J2: jumper
3 to 5.
SW1: set
2 closed.
J2: jumper
3 to 5.
SW1: set
2 and 4 closed.
J2: jumper
3 to 5.
SW1: use
default settings.
J2: no jumpers
TM-XROP-01XN.book Page 69 Monday, March 22, 2004 4:10 PM