Quality assurance document for the CT 200 measuring
instrument
If you have any questions do not hesitate to contact us at any time.
Mail: m.kloeckner@xonox-tec.com // Phone: +49 6441 21049-13
Check of mechanical alignment
1) Place 10mm gauge block on the lens transducer and tighten
the chuck.
2) Switch on the digital dial gauge and hold the SET / ZERO button for two seconds. Now
in the display appears +10.0000 mm and P flashes. Pressing the Set button to set this
value (+10.0000 mm).
3) Operate the left switch (the measuring tips touch the 10mm gauge block). It is important
that the upper measuring tip first touch the gauge block. Then, press the SET button
on the remote controller. The gauge is now calibrated for +10.0000 mm.