Note: UT is the A.C. mains voltage prior to application of the test level.
Table 3: Guidance & Declaration - electromagnetic immunity concerning
Conducted RF & Radiated RF
Guidance & Declaration - Electromagnetic immunity
The M4 are intended for use in the electromagnetic environment
specified below. The customer or the user of the M4 should assure
that they are used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment -
guidance
Conducted RF
lEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM
bands
3 Vrms
150
kHz to 80 MHz
6 Vrms in ISM &
amateur radio
bands
Portable RF
communications
equipment
(including
peripherals such
as antenna cables
and external
antennas) should
be used no closer
than 30 cm (12
inches) to any part
of the
M4, including
cables specified
by the
manufacturer.
Otherwise,
degradation of the
Radiated RF
lEC 61000-4-3
10 V/m
80 MHz to 2.7
GHz
385MHz-
5785MHz Test
specifications
for
ENCLOSURE
PORT
20V/m
26MHz to 25GHz
10 V/m
80 MHz to 2.7
GHz
385MHz-
5785MHz Test
specifications for
ENCLOSURE