OxiTop
®
(-i/-IDS) What to do if...
ba77059e01 03/2019 27
6 What to do if...
Displayed measured
value too low
Displayed measured
value too high
Signal LED flashes
red
Measuring head
does not react to
keystroke
Cause Remedy
The measuring equipment is
untight
Insufficient sample preparation,
lacking sample conservation.
The temperature of the sample
had not sufficiently adapted (<
15°C).
Too few NaOH tablets
Tighten the measurement equip-
ment
Carry out suitable sample prepara-
tion, sample conservation
If possible, repeat the measurment.
Use the recommended amount of
fresh NaOH tablets
Cause Remedy
The measuring range selected is
too small.
Missing or lacking nitrification
inhibitor (ATH)
With very high values
(> 2000 mg/l):
Dilute the sample and start a new
measurement.
Use enough nitrification inhibitor
(ATH)
Cause Remedy
Allowed pressure exceeded
(OFL is displayed, only with
anaerobic degradation tests)
Vent the sample bottle
Battery nearly discharged
(the symbol for low battery is dis-
played)
The following measures are taken auto-
matically during a measurement to
avoid data loss:
Radio connection to the meter (only
for OxiTop
®
-IDS): Off
Signal LED (green, blue): OFF
The LED flashes red only.
Display illumination: Off
Switch-off time: 1 min
The measurement can still be contin-
ued for 7 days after the low battery dis-
play appears.
Cause Remedy
Operating condition undefined Press the
<M> and <
> keys
simultaneously (reset)