5 - 2 Communication Data Table
20
Chapter 5. COMMUNICATION DATA TABLE
The enabling conditions for the address and R/W (Read/Write) of each data are specified in the following table:
The meaning of R/W column marks:
Possible
Impossible
Device related data
Note
For RD and WD commands, the data is required four hexadecimal digits.
Item Data range RAM EEPROM Remarks
Address
RW
Address
RW
Decimal Decimal
(Hexadecimal) (Hexadecimal)
Gas type 0: User setting 1001 4001 ✕✕ Gas type change is
1: Nitrogen/Air (03E9) (0FA1) possible by function
2: Oxygen setup (Address 5018)
3: Argon
4: Carbon dioxide
5: Natural gas 13A
(Heating value: 46MJ/m
3
)
6: Propane
7: Methane
8: Butane
9: Hydrogen
10:Helium
11:Natural gas 13A
(Heating value: 45MJ/m
3
)
Full- scale flow Depend on flow rate 1002 4002 ✕✕ The decimal point
rate range (03EA) (0FA2) is removed.
Decimal point 0: No decimal point 1003
4003 ✕✕
position of 1: XXXX. (03EB) (0FA3)
instantaneous 2: XXX.X
flow rate 3: XX.XX
4: X.XXX
Decimal point 0: No decimal point 1004
4004 ✕✕
position of 1: XXXXXXXX. (03EC) (0FA4)
integrated 2: XXXXXXX.X
flow rate 3: XXXXXX.XX
4: XXXXX.XXX
Instantaneous 0: mL/min 1005
4005 ✕✕
flow rate unit 1: L/min (03ED) (0FA5)
Integrated 0: L 1006
4006 ✕✕
flow rate unit 1: m
3
(03EE) (0FA6)