3-22
IM AQ6375E-02EN
CAUTION
Before purge with nitrogen, ensure that there is adequate ventilation in the room.
The instrument warranty will be void if damage is caused by the use of inappropriate
nitrogen and purge gas.
French
ATTENTION
Avant toute purge à l’azote, vérifiez que la pièce est correctement ventilée.
La garantie de l’instrument est nulle si les dommages sont causés par l’utilisation d’azote
ou d’air sec inadaptés.
Purge gas
Use ultra-high purity liquid nitrogen which must be 99.9999% pure grade with the heat exchanger is
recommended. The gas temperature must be 23±3°C.
If compressed nitrogen must be used, the gas must be dry, oil-free and uncontaminated. Never use
compressed nitrogen from a supplier who uses oil or water in the compression process.
Maximum rating
Pressure and flow rate of purge gas must be lower than the following maximum rating.
Pressure: 1.5 psig (0.01 MPaG)
Flow rate: 25 SCFH (12 L/min)
Purge Gas Temperature Specifications
Gas temperature: 23±3°C
Note
A pressure regulator and precision flowmeter must be used.
Aprecision flowmeter with an adjustable needle valve is recommended.
3.6 Purge