5-24 <5. Functions of CQ1 Software>
IM 80J01A01-01E
Flat Field Correction
Correct peripheral image darkness caused by shading of the
microscope. (Default: checked)
Chromatic Aberration
Correct chromatic aberration caused by difference of wavelength.
(Default: checked)
Laser Enhancement
Check to enhance excitation power. If checked, value of excitation
power turns red.
Normally excitation power is calibrated by power irradiated
on sample. If check this box, the calibration is disabled and
excitation laser is output by non-calibrated laser power.