9-20
IM CW120-E
9.6.3 Wiring for Terminals
Wiring for External Control Input/Output Terminals
With a flatblade screwdriver or the like, push in the rectangular dip
beneath each terminal hole and insert a peeled signal wire. Releasing the
screwdriver grips the wire.
signal wire
Flatblade screwdriver or the like
NOTE
Do not apply a voltage exceeding the allowable voltage range (–0.5 to 5.5 V) to the control
signal input terminal. The input circuit may be damaged.
Do not confuse the input terminal with the output terminal.
Applicable Wires for Signal Wiring
Recommended wires:
Solid AWG 18 wires (1.0 mm diameter) or stranded 0.75 mm
2
wires
Usable wires:
Solid AWG 26 to AWG 18 (0.4 to 1.0 mm diameter) wires
Stranded AWG 22 to AWG 20 (0.3 to 0.75 mm
2
diameter) wires with
each strand thicker than 0.18 mm diameter
Standard peeling length: 10 mm