14-10
IM CW120-E
(2) 96032 Clamp-on Current Probe
Measurement range
Output voltage
Accuracy
Temperature coefficient
Maximum allowed current
Output impedance
External magnetic field effects
Conductor position effects
Circuit voltage
Withstanding voltage
Measurable conductor
diameter
Operating temperature and
humidity ranges
Storage temperature and
humidity ranges
Applicable locations
Dimensions (W × H × D)
Weight
Output terminals
Output cable length
Standard accessories
0 to 1000 Arms AC (1414 Apk) for 5 minutes
0 to 0.25 Vrms AC (0.25 mV/A)
±1.0% of rdg ±0.2 mV (45 Hz to 66 Hz)
±1.0° (for 50 A or higher, 45 Hz to 66 Hz)
Under the conditions of the temperature at 23°C ±5°C, humidity at 35
to 75% RH, with sine wave input
0.05% of full scale/°C in ranges of 5°C to 40°C
1000 Arms AC for 5 minutes, 700 Arms AC continuous (45 Hz to 66 kHz)
Approximately 100 (maximum)
0.5 A equivalent or less (at 400 A/m, 50/60 Hz)
±0.5% (at 200 to 1000 A, 45 Hz to 66 kHz)
600 Vrms AC maximum
2.2 kVrms AC for 1 minute (core-to-casing, and core-to-output)
65 mm diameter maximum, 65 × 75 mm busbar maximum
5° to 40°C, 35 to 80% RH (no condensation)
–20° to 60°C, 90% RH or less (no condensation)
Indoors, altitude of 2000 m or lower
Approximately 100 × 172.5 × 32 mm
Approximately 500 g
Banana plugs (safety terminals)
Approximately 3 meters
One User’s Manual, ring markers (4 colors × 2 each, L4007MG)
Amplitude
Phase
Item 96032 (Max. 1000-A Range)
Dimensions of 96032 Clamp-on Current Probe
32
(1.3)
24.7
(1.0)
100 (3.9)
66 (2.6)
71 (2.8)
172.5 (6.8)
Unit: mm
(approx. inches)
14.2 Specifications of Clamp-on Current Probes