App 2-22
Appendix 2. Communication Commands
IM CW120-E
Communication Execution Errors (Codes 200 to 299)
200
221
222
223
224
241
Execution error
Setting conflict
Data out of range
Too much data
Illegal parameter value
Hardware missing
Code Message Countermeasure
The specified command cannot be carried out.
Check the setting with other related settings.
Check the setting range.
Check the byte length of the parameter.
Check the setting range.
Check the options equipped.
Query Errors (Codes 400 to 499)
410
420
430
440
Query INTERRUPTED
Query UNTERMINATED
Query DEADLOCKED
Query UNTRMINATED
after indefinite response
Code Message Countermeasure
Check the transmission-reception sequence.
Check the transmission-reception sequence.
Limit the length of each program message within 1024
bytes including the <PMT>.
Do not code a query following “*IDN?”.