4-1
IM CW120-E
Connecting Power Supply and Turning Power On/Off
4
Chapter 4. Connecting Power Supply and Turning Power On/Off
4.1 When Using AC Power Supply
Before Connecting Power Supply
There is a danger of electric shock or damage to the meter. Observe the
following warning notes.
WARNING
Use only the dedicated power cord supplied by the manufacturer.
Before connecting the power cord, make sure the power-source voltage matches the rated
supply voltage.
Before connecting the power cord, also make sure the power switch of the CW120 or CW121 is
turned off.
If the CW120 or CW121 will not be used for a prolonged period, disconnect the power cord
from the outlet.
Do not place any load on the power cord or allow the power cord to come into accidental
contact with a heat source.
Be sure to hold the plug of the power cord, rather than holding and pulling the cord itself,
when disconnecting it from the outlet.
Procedure for Connecting Power Cord
Follow the steps below to connect the power cord.
1. Make sure the power switch of the CW120/121 is turned off.
2. Plug one end of the power cord supplied with the CW120/121 into the
power supply jack on the side of the CW120/121.
3. Plug the other end of the power cord into a power outlet that meets the
requirements below.
Required Power Ratings
Rated supply voltage 100 to 240 V AC
Allowable supply voltage range 90 to 264 V AC
Rated power supply frequency 50 or 60 Hz
Allowable supply frequency range 48 to 63 Hz
Maximum power consumption 8 VA (at 240 V AC)