18-7
IM 701210-06E
Troubleshooting, Maintenance, and Inspection
18
18.2 Messages and Corrective Actions
Code Message and Corrective Action Section
887 A sampling speed was too fast, and it could be folded, and a real-time queue was needed. 7.7
Slow down a sampling speed.
888 A sampling speed was too fast, and it could be folded, and a real-time buffer was needed. 7.7
Slow down a sampling speed.
889 Cannot set On this module. 6.17
890 Settings can not be entered for channels on which no strain module is mounted.
891 Cannot be changed when Variable is turned ON. Chapter 15
892 A DSP channel cannot be specified for this trigger type. Chapter 15
893 Cannot be set if Operation is set to something other than KnockFlt. Chapter 15
894 Cannot set this module when Operation is set to KnockFlt. Chapter 15
895 Cannot be set when the trace is set to a frequency module. Chapter 15
896 Cannot be set when the trace is set to a DSP channel. Chapter 15
897 The capture window cannot be changed while the dual capture is in progress, and while the measuring 7.6
is in progress.
898 Cannot be set while the capture window is opened. 7.6
System Operation Errors (900 to 999)
Code Message and Corrective Action Section
901 Failed to backup setup data. Will initialize. Backup battery may be low. 7.10
902 System RAM failure. Maintenance service is required.
903 System ROM failure. Maintenance service is required.
906 Fan stopped. Power off immediately. Maintenance service is required.
907 Backup battery is flat. Maintenance service is required to replace the back-up battery. 3.4
908 Internal temperature is too high. Power off immediately. Maintenance service is required. 3.4
909 Illegal SUM value. Maintenance service is required. 3.4
910 RAM read/write error. Maintenance service is required.
911 Memory bus error. Maintenance service is required.
912 Communication driver error.
914 Time out occurs in Communication.
915 EEPROM read error. EEPROM may be damaged. Maintenance service is required.
916 EEPROM write error. EEPROM may be damaged. Maintenance service is required.
917 No module installed. Install the module. 3.3
918 Turn ON the internal hard disk motor. 13.2
919 The current module installation condition and the setup data are inconsistent.
Will initialize.
922 Cannot back up the acquisition memory. Memory will be initialized. 7.10
The battery for acquisition memory backup may be flat.
928 Battery for waveform memory is flat. Waveform memory was initialized. 7.10
929 MS bus error occurred.
930 Slot 1 EEPROM error.
931 Slot 2 EEPROM error.
932 Slot 3 EEPROM error.
933 Slot 4 EEPROM error.
934 Slot 5 EEPROM error.
935 Slot 6 EEPROM error.
936 Slot 7 EEPROM error.
937 Slot 8 EEPROM error.
938 Key protect is enabled. To release the protection, press the PROTECT key. 17.6
939 This firmware is for the DL750P. It cannot be used on the DL750. 17.6
940 The USB device’s power consumption exceeded the capacity of the USB hub. 19.8