EasyManua.ls Logo

YOKOGAWA EJA210E - Page 11

YOKOGAWA EJA210E
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<2. Handling Cautions>
10
IM01C25A01-01E
c. FM Intrinsically Safe/Nonincendive for
Fieldbus Type for EJX90A
EJXmultivariabletransmitterwithoptionalcode
/FS15areapplicableforuseinhazardouslocations.
Note1.
• ApplicableStandard:
Class3600:2011,Class3610:2010,
Class3611:2004,Class3810:2005
ANSI/ISA-60079-0:2009,ANSI/ISA-60079-11:2009,
ANSI/ISA-60079-27:2006
NEMA250:1991
• Rating
IntrinsicallySafewithEntity/FISCOforClassI,II,III
Division1GroupsA,B,C,D,E,F,GT4
ClassIZone0AExiaIICT4
Nonincendiveeldwiring/FNICOforClassI,II
Division2GroupsA,B,C,D,F,GT4
ClassIIIDivision1T4
ClassI,Zone2GroupIICT4
• Ambienttemperature:–40to60°C
• Enclosure:Type4X
• ElectricalConnection:1/2NPTfemale,M20
female
Note2. Installation
InstallationshouldbeinaccordancewithControl
DrawingIFM026-A12.

Related product manuals