<2. Handling Cautions>
18
IM01C25A01-01E
Electrical Data
• WhencombinedwithTrapezoidalandRectangular
outputcharacteristicFISCOmodelIICbarrier
Ui=17.5V,Ii=380mA,Pi=5.32W,
Ci=3.52nF,Li=0μH
• WhencombinedwithLinearcharacteristicbarrier
Ui=24V,Ii=250mA,Pi=1.2W,
Ci=3.52nF,Li=0μH
• WhencombinedwithTrapezoidalorRectangular
outputcharacteristicFISCOmodelIIBbarrier
Ui=17.5V,Ii=460mA,Pi=5.32W,
Ci=3.52nF,Li=0μH
Note3. Installation
• Allwiringshallcomplywithlocalinstallation
requirements.(Refertotheinstallationdiagram)
Note4. MaintenanceandRepair
• Theinstrumentmodicationorpartsreplacement
byotherthanauthorizedrepresentativeof
YokogawaElectricCorporationisprohibitedand
willvoidDEKRAIntrinsicallysafeCertication.
Note5. SpecialConditionsforSafeUse
WARNING
• InthecasewheretheenclosureofthePressure
Transmitterismadeofaluminium,ifitismounted
inanareawheretheuseofcategory1G
apparatusisrequired,itmustbeinstalledsuch,
thatevenintheeventofrareincidents,ignition
sourcesduetoimpactandfrictionsparksare
excluded.
• Electrostaticchargemaycauseanexplosion
hazard.Avoidanyactionsthatcausethe
generationofelectrostaticcharge,suchasrubbing
withadryclothoncoatingfaceoftheproduct.
• InthecasewheretheenclosureofthePressure
Transmitterismadeofaluminum,ifitismountedin
anareawheretheuseofcategory2Dapparatusis
required,itshallbeinstalledinsuchawaythatthe
riskfromelectrostaticdischargesandpropagating
brushdischargescausedbyrapidowofdustis
avoided.
• TosatisfyIP66orIP67,applywaterproofglandsto
theelectricalconnectionport.
• Whenthelightningprotectoroptionisspecied,
theapparatusisnotcapableofwithstandingthe
500VinsulationtestrequiredbyEN60079-11.
Thismustbetakenintoaccountwheninstalling
theapparatus.
Note6. InstallationInstructions
Non-Hazardous Location
Hazardous Location
F0231.ai
Terminator
Safety Barrier
Transmitter
Pressure
Transmitter
Transmitter
+
+
+
Terminator
+
+
[Installation Diagram]
SUPPLY
• Intherating1(*1),theoutputcurrentofthebarrier
mustbelimitedbyaresistor‘Ra’suchthatIo=Uo/
Ra.
• Intherating2(*2),theoutputofthebarriermustbe
thecharacteristicsofthetrapezoidortherectangle
andthistransmittercanbeconnectedtoFieldbus
equipmentwhichareinaccordingtotheFISCO
model.
• Theterminatorsmaybebuiltinbyabarrier.
• Morethanonetransmittermaybeconnectedto
thepowersupplyline.
• Theterminatorandthesafetybarriershallbe
certied.
Electricaldata:
MaximumInputVoltageUi:24V
*1:
Rating 1
MaximumInputCurrentIi:250mA
MaximumInputPowerPi:1.2W
MaximumInternalCapacitanceCi:3.52nF
MaximumInternalInductanceLi:0μH
or
MaximumInputVoltageUi:17.5V
*2:
Rating 2
MaximumInputCurrentIi:380mA
MaximumInputPowerPi:5.32W
MaximumInternalCapacitanceCi:3.52nF
MaximumInternalInductanceLi:0μH
or
MaximumInputVoltageUi:17.5V
MaximumInputCurrentIi:460mA
MaximumInputPowerPi:5.32W
MaximumInternalCapacitanceCi:3.52nF
MaximumInternalInductanceLi:0μH