EasyManua.ls Logo

YOKOGAWA EJX Series - 2.4 IECEx Certification

YOKOGAWA EJX Series
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<2.HandlingCautions>
14
IM01C25A01-01E
(4) Operation
WARNING
• OPENCIRCUITBEFOREREMOVINGCOVER.
INSTALLINACCORDANCEWITHTHISUSER’S
MANUAL
• Takecarenottogeneratemechanicalsparking
whenaccesstotheinstrumentandperipheral
devicesinahazardouslocation.
(5) MaintenanceandRepair
WARNING
Theinstrumentmodicationorpartsreplacementby
otherthananauthorizedRepresentativeofYokogawa
ElectricCorporationisprohibitedandwillvoidthe
certication.
(6) NamePlate
F0215.ai
Name plate
Tag plate for flameproof type
Tag plate for intrinsically safe Ex ia
WARNING
D
D
WARNING
No. KEMA 07ATEX0109 X
Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db
Enlcosure : IP66/IP67
TEMP. CLASS T6 T5 T4
MAX PROCESS TEMP.(Tp.) 85 100 120 °C
Tamb. -50 to 75 80 75 °C
T85°C(Tamb.:-30(-15) to 75°C, Tp.:85°C)(for Dust)
No. DEKRA 11ATEX 0228 X
Ex ia IIC T4 Ga Ta: -50 TO 60°C
Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) TO 60°C
IP66/IP67
MAX. PROCESS TEMP.(Tp.) 120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Ui=30V, Ii=200mA , Pi=0.9W, Ci=27.6nF, Li=0µH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
- SEE USER’S MANUAL
AFTER DE-ENERGIZING, DELAY 5 MINUTES BEFORE
OPENING.
WHEN THE AMBIENT TEMP. 65°C, USE THE
HEAT-RESISTING CABLE & CABLE GLAND ≥ 90°C
POTENTIAL ELECTROSTATIC CHARGING HAZARD
Tag plate for intrinsically safe type
(for EJX910A and EJX930A)
D
No. KEMA 06ATEX0037 X
EEx ia IIC T4
IP66 and IP67
Tamb. -50(-15) to 60°C MIN Tamb.for DUST -40(-15°C)
MAX PROCESS TEMP.(Tp) 120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Supply/Pulse circuit
Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH
Sensor circuit
Uo=30 V, Io=95.4 mA, Po=468 mW, Co=11 nF, Lo=3.9 mH
Tag plate for intrinsically safe Ex ic
WARNING
Ex ic IIC T4 Gc
IP66
Tamb -30(-15) TO 60°C
MAX. PROCESS TEMP. 120°C
Ui=30V, Ci=27.6nF, Li=0µH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
- SEE USER’S MANUAL
*3
*3
*3
F0215.ai
Name plate
Tag plate for flameproof type
Tag plate for intrinsically safe Ex ia
WARNING
D
D
WARNING
No. KEMA 07ATEX0109 X
Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db
Enlcosure : IP66/IP67
TEMP. CLASS T6 T5 T4
MAX PROCESS TEMP.(Tp.) 85 100 120 °C
Tamb. -50 to 75 80 75 °C
T85°C(Tamb.:-30(-15) to 75°C, Tp.:85°C)(for Dust)
No. DEKRA 11ATEX 0228 X
Ex ia IIC T4 Ga Ta: -50 TO 60°C
Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) TO 60°C
IP66/IP67
MAX. PROCESS TEMP.(Tp.) 120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Ui=30V, Ii=200mA , Pi=0.9W, Ci=27.6nF, Li=0µH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
- SEE USER’S MANUAL
AFTER DE-ENERGIZING, DELAY 5 MINUTES BEFORE
OPENING.
WHEN THE AMBIENT TEMP. 65°C, USE THE
HEAT-RESISTING CABLE & CABLE GLAND 90°C
POTENTIAL ELECTROSTATIC CHARGING HAZARD
Tag plate for intrinsically safe type
(for EJX910A and EJX930A)
D
No. KEMA 06ATEX0037 X
EEx ia IIC T4
IP66 and IP67
Tamb. -50(-15) to 60°C MIN Tamb.for DUST -40(-15°C)
MAX PROCESS TEMP.(Tp) 120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Supply/Pulse circuit
Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH
Sensor circuit
Uo=30 V, Io=95.4 mA, Po=468 mW, Co=11 nF, Lo=3.9 mH
Tag plate for intrinsically safe Ex ic
WARNING
Ex ic IIC T4 Gc
IP66
Tamb -30(-15) TO 60°C
MAX. PROCESS TEMP. 120°C
Ui=30V, Ci=27.6nF, Li=0µH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
- SEE USER’S MANUAL
*3
*3
*3
MODEL:Speciedmodelcode.
STYLE:Stylecode.
SUFFIX:Speciedsufxcode.
SUPPLY:Supplyvoltage.
OUTPUT:Outputsignal.
MWP:Maximumworkingpressure.
CALRNG:Speciedcalibrationrange.
NO.:Serialnumberandyearofproduction
*1
.
TOKYO180-8750JAPAN:
Themanufacturernameandtheaddress
*2
.
*1:Therstnumberinthesecondblockof“NO.”column
isthelastonenumberoftheproductionyear.
second block
NO. 91K819857 1
32 7
The year 2011
*2:“180-8750”isazipcodewhichrepresentsthe
followingaddress.
2-9-32Nakacho,Musashino-shi,TokyoJapan
*3:TheidenticationnumberofNotiedBody.
2.4.4 IECExCertication
ModelEJXSeriespressuretransmitterswith
optionalcode/SU2canbeselectedthetypeof
protection(IECExIntrinsicallySafe/typenor
ameproof)foruseinhazardouslocations.
Note1. Fortheinstallationofthistransmitter,oncea
particulartypeofprotectionisselected,any
othertypeofprotectioncannotbeused.The
installationmustbeinaccordancewiththe
descriptionaboutthetypeofprotectioninthis
instructionmanual.
Note2. Inordertoavoidconfusion,unnecessary
markingiscrossedoutonthelabelother
thantheselectedtypeofprotectionwhenthe
transmitterisinstalled.
a. IECExIntrinsicallySafeType/typen
CautionforIECExIntrinsicallysafeandtypen.
Note1. ModelEJXSeriesdifferential,gauge,and
absolutepressuretransmitterswithoptional
code/SU2areapplicableforuseinhazardous
locations
• No.IECExCSA05.0005
• ApplicableStandard:IEC60079-0:2000,
IEC60079-11:1999,IEC60079-15:2001

Related product manuals