<2. Handling Cautions>
10
IM01C25A01-01E
c. FM Intrinsically Safe/Nonincendive for
Fieldbus Type for EJX90A
EJXmultivariabletransmitterwithoptionalcode
/FS15areapplicableforuseinhazardouslocations.
Note1.
• ApplicableStandard:
Class3600:2011,Class3610:2010,
Class3611:2004,Class3810:2005
ANSI/ISA-60079-0:2009,ANSI/ISA-60079-11:2009,
ANSI/ISA-60079-27:2006
NEMA250:1991
• Rating
IntrinsicallySafewithEntity/FISCOforClassI,II,III
Division1GroupsA,B,C,D,E,F,GT4
ClassIZone0AExiaIICT4
Nonincendiveeldwiring/FNICOforClassI,II
Division2GroupsA,B,C,D,F,GT4
ClassIIIDivision1T4
ClassI,Zone2GroupIICT4
• Ambienttemperature:–40to60°C
• Enclosure:Type4X
• ElectricalConnection:1/2NPTfemale,M20
female
Note2. Installation
InstallationshouldbeinaccordancewithControl
DrawingIFM026-A12.