<2. Handling Cautions>
23
IM01C25A01-01E
(5) Maintenance and Repair
WARNING
Theinstrumentmodicationorpartsreplacementby
otherthananauthorizedRepresentativeofYokogawa
ElectricCorporationisprohibitedandwillvoidthe
certication.
(6) Name Plate
F0215-1.ai
Name plate
Tag plate for flameproof type
D
WARNING
No. KEMA 07ATEX0109 X
Ex d IIC T6...T4 Gb, Ex tb IIIC T85°C Db
Enlcosure : IP66/IP67
TEMP. CLASS T6 T5 T4
PROCESS TEMP.(Tp.) -50 TO 85 100 120 °C
Tamb. (for GAS) -50 TO 75 80 75 °C
T85°C(Tamb.: -30(-15) TO 75°C, Tp.:-30(-15) TO 85°C)(for Dust)
AFTER DE-ENERGIZING, DELAY 5 MINUTES BEFORE
OPENING.
WHEN THE AMBIENT TEMP. ≥ 65°C, USE THE
HEAT-RESISTING CABLE & CABLE GLAND ≥ 90°C
POTENTIAL ELECTROSTATIC CHARGING HAZARD -SEE THE USER’S MANUAL
*3
F0215-2.ai
Tag plate for intrinsically safe Ex ia for HART/BRAIN
WARNING
D
No. DEKRA 11ATEX 0228 X
Ex ia IIC T4 Ga Ta: -50 TO 60°C
Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) TO 60°C
IP66/IP67
MAX. PROCESS TEMP.(Tp.) 120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Ui=30V, Ii=200mA , Pi=0.9W, Ci=27.6nF, Li=0µH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
- SEE USER’S MANUAL
[Except for EJX90A]
Tag plate for intrinsically safe Ex ic for HART/BRAIN
WARNING
Ex ic IIC T4 Gc
IP66
Tamb -30(-15) TO 60°C
MAX. PROCESS TEMP. 120°C
Ui=30V, Ci=27.6nF, Li=0µH
POTENTIAL ELECTROSTATIC
CHARGING HAZARD
- SEE USER’S MANUAL
*3
F0215-3.ai
Tag plate for intrinsically safe Ex ia for Fieldbus
No. KEMA 06ATEX0278 X
Ex ia IIC/IIB T4 Ga Ta: -40 to 60°C
Ex ia IIIC T85°C T100°C T120°C Db Ta: -30(-15) to 60°C
MAX PROCESS TEMP.(Tp.): 120°C
T85°C (Tp.: 80°C), T100°C (Tp.: 100°C), T120°C (Tp.: 120°C)
ENCLOSURE: IP66/IP67
Supply circuit
FISCO field device(IIC/IIB)
Entity Parameter Ui=24V, Ii=250mA, Pi=1.2W, Ci=3.52nF, Li=0µH
Sensor circuit
Uo=7.63V, Io=3.85mA, Po=0.008W, Co=4.8uF, Lo=100mH
POTENTIAL ELECTROSTATIC CHARGING HAZARD-SEE USER'S MANUAL.
*3
[EJX90A]
Tag plate for intrinsically safe Ex ia for Fieldbus
WARNING
No. KEMA 04ATEX1116 X
Ex ia IIC/IIB T4 Ga Ta:-40 to 60°C
Ex ia IIIC T85°C T100°C T120°C Db Ta:-30(-15) to 60°C
MAX PROCESS TEMP.(Tp.):120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Enclosure: IP66/IP67
FISCO field device(IIC/IIB)
Entity parameters Ui=24V, Ii=250mA, Pi=1.2W, Ci=3.52nF, Li=0μH
WARNING
POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE USER'S MANUAL.
D
Tag plate for intrinsically safe Ex ic for Fieldbus
Ex ic IIC T4 Gc
IP66
Tamb -30(-15) TO 60°C
MAX. PROCESS TEMP. 120°C
Ui=32V, Ci=3.52nF, Li=0µH
POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE USER'S MANUAL.
*3
MODEL:Speciedmodelcode.
STYLE:Stylecode.
SUFFIX:Speciedsufxcode.
SUPPLY:Supplyvoltage.
OUTPUT:Outputsignal.
MWP:Maximumworkingpressure.
CALRNG:Speciedcalibrationrange.
NO.:Serialnumberandyearofproduction
*1
.
TOKYO180-8750JAPAN:
Themanufacturernameandtheaddress
*2
.
*1:Therstnumberinthesecondblockof“NO.”column
isthelastonenumberoftheproductionyear.
second block
NO. 91K819857 1
32 7
The year 2011
*2:“180-8750”isazipcodewhichrepresentsthe
following address.
2-9-32Nakacho,Musashino-shi,TokyoJapan
*3:TheidenticationnumberofNotiedBody.