EasyManua.ls Logo

YOKOGAWA EJX110A - Page 26

YOKOGAWA EJX110A
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<2. Handling Cautions>
25
IM01C25A01-01E
Note2. ElectricalData
MaximumInputVoltage(Ui)=30V
MaximumInternalCapacitance(Ci)=27.6nF
MaximumInternalInductance(Li)=0μH
Note3. Installation
• Thepressuretransmitterisallowedtobeinstalled
in“nL”systems,onconditionthattheoutput
parametersof“nL”source(associatedenergy-
limitedapparatus)aresuitabletotheabove
mentionedinputparametersofthepressure
transmitterandthecableparameters.
• Cableglands,adaptersand/orblankingelements
shallbeofEx“n”,Ex“e”orEx“d”andshallbe
installedsoastomaintainthespecieddegreeof
protectionoftheequipment.
• Theinstrumentmodicationorpartsreplacement
byotherthanauthorizedrepresentativeof
YokogawaElectricCorporationisprohibitedand
willvoidIECExcertication.
F0237.ai
EJX/EJA-E Series
Pressure Transmitters
Associated apparatus
Supply
Hazardous Location Nonhazardous Location
+
+
[Ex ic]
Note4. SpecicConditionofUse
WARNING
• Electrostaticchargemaycauseanexplosion
hazard.Avoidanyactionsthatcausethe
generationofelectrostaticcharge,suchasrubbing
withadryclothoncoatingfaceoftheproduct.
• Theapparatusisnotcapableofdielectricstrength
testsrequiredbyIEC60079-11.
Thismustbetakenintoaccountwheninstalling
theapparatus.
c. IECEx Intrinsically Safe Ex ia for Fieldbus
Type (Except for EJX90A)
CautionforIECExIntrinsicallysafeExia.
Note1. EJX/EJA-ESeriespressuretransmitterswith
optionalcode/SS26areapplicableforusein
hazardouslocations.
• No.IECExDEK12.0016X
• ApplicableStandard:IEC60079-0:2011,
IEC60079-11:2011,IEC60079-26:2006
• ExiaIIC/IIBT4Ga
• AmbientTemperature:–40to60°C
• Max.ProcessTemp.:120°C
Note2. ElectricalData
• Intrinsicallysaferatingsareasfollows:
[Entity]
MaximumInputVoltage(Ui)=24V
MaximumInputCurrent(Ii)=250mA
MaximumInputPower(Pi)=1.2W
MaximumInternalCapacitance(Ci)=3.52nF
MaximumInternalInductance(Li)=0μH
[FISCOIIC]
MaximumInputVoltage(Ui)=17.5V
MaximumInputCurrent(Ii)=380mA
MaximumInputPower(Pi)=5.32W
MaximumInternalCapacitance(Ci)=3.52nF
MaximumInternalInductance(Li)=0μH
[FISCOIIB]
MaximumInputVoltage(Ui)=17.5V
MaximumInputCurrent(Ii)=460mA
MaximumInputPower(Pi)=5.32W
MaximumInternalCapacitance(Ci)=3.52nF
MaximumInternalInductance(Li)=0μH
Note3. Installation
Non-Hazardous Location
Hazardous Location
F0238.ai
Terminator
Safety Barrier
Field Instruments
Pressure
Transmitter
Field Instruments
+
+
+
Terminator
+
+
[Ex ia]
• Intherating1(*1),theoutputcurrentofthebarrier
mustbelimitedbyaresistor‘Ra’suchthatIo=Uo/
Ra.
• Intherating2(*2),theoutputofthebarriermustbe
thecharacteristicsofthetrapezoidortherectangle
andthistransmittercanbeconnectedtoFieldbus
equipmentwhichareinaccordingtotheFISCO
model.
• Theterminatorsmaybebuiltinbyabarrier.
• Morethanonetransmittermaybeconnectedto
thepowersupplyline.
• Theterminatorandthesafetybarriershallbe
certied.

Related product manuals