89
IM FG410-01EN
d) 3rdpageofsettingscreen:Screenforcomplementarysettingsforsweep
The figure is an example that shows frequency selection as the sweep type.
Indicates that the
3rd page is
displayed.
Sweep control with
multi-I/O connector
Stop level of gated
singleshot sweep
Gated sweep
oscillation stop unit
Sync output
Two soft key stages
Press the NEXT
key to switch to
the 1st page.
Stop level [StpLvl]
This is the signal level when gated single-shot sweep oscillation is stopped. It can be set either to
[Off] or [On] for setting the level. Normally, [Off] is set.
p.93
External control with multi-I/O connector [ExtCtrl]
Enables/disables external control with the multi-I/O connector.
p.98
Oscillation stop unit for gated single-shot sweep [OscStop]
This is the oscillation stop unit for gated single-shot sweep. The unit can be selected from 1 cycle
and 1/2 cycle. Normally, the unit is set to 1 cycle.
p.93
Sync output [SyncOut]
This is the output signal from the sync/sub-output terminal. Waveform reference phase sync, sweep
sync, sweep marker, and sweep X drive can be selected from.
p.96
4.8 Sweep Setting and Manipulation