EasyManua.ls Logo

YOKOGAWA FG420 - Page 108

YOKOGAWA FG420
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
94
IM FG410-01EN
4. BASIC OPERATION
Gatedsingle-shotone-waysweep
Sweep value
Frequency sweep example
Start value
Stop value
Sweep time
Trigger
Start value
Gatedsingle-shotshuttlesweep
Sweep value
Frequency sweep example
Start value
Stop value
Sweep time Sweep time
Trigger
Check
In the case of phase sweep, the start phase setting is the oscillation start phase, and the
stop phase setting is the oscillation stop phase.
k) Tosettriggerconditionforsingle-shotsweepandgatedsingle-shotsweep
The trigger can be selected from internal trigger oscillator, external signal, manual trigger
manipulation, and remote trigger manipulation.
Upon receipt of a trigger, the TRIG’D lamp next to the MAN TRIG key lights.
The trigger condition is set with [Trig] (trigger) on the 2nd page of the setting screen.
Triggersourcesetting
The trigger source can be chosen from [Int] (internal) and [Ext] (external).
When the trigger source is [Int] (internal), the trigger period can be set.
When the trigger source is [Ext] (external), the trigger polarity can be set.
If the trigger source is set to [Ext] (external), a TTL level trigger signal is input to the external
trigger input terminal (TRIG IN).
Sweeping can be started and stopped independently of the trigger source setting through input of
a logic signal to the multi-I/O connector.
p.98
Usage of manualtriggerandremotetrigger
Manual trigger manipulation and remote trigger manipulation are always effective regardless of
the trigger source setting.
The [Start] soft key and the
MAN TRIG
MAN TRIG key can be used for manual trigger manipulation.

Table of Contents