103
IM FG410-01EN
4.8.6 Settingamplitudesweep
For the manipulation methods that are common with the sweep setting screen, refer to p.87 and
p.90 .
a) Amplitudesweepexample
The following is an example of continuous, linear shuttle sweep.
Shuttlesweep
Sweep value
Output signal
Start value
Stop value
Sweep time Sweep time
b) Selectingamplitudesweep
When [Mode] (oscillation mode) is set
to [Sweep] (sweep), set [Type] (sweep
type) on the 2nd page of the setting
screen to [Amptd] (amplitude).
c) Waveformsforwhichamplitudesweepisnotpossible
Amplitude sweep is not possible for DC.
d) Settingitemsrequiredforamplitudesweep
Set the following items on the 2nd page of the setting screen. The amplitude settings on the 1st page
of the setting screen are disabled.
• Start amplitude [Start]
• Stop amplitude [Stop]
• Sweep time [Time]
This is the transition time from the start amplitude to the stop amplitude.
p.91
• Sweep mode [SwpMode]
Continuous, single-shot, and gated single-shot can be chosen from.
p.91
• Sweep function [SwpFctn]
One-way/shuttle can be chosen from.
p.90
Setting with [Center] (center amplitude) and [Span] (span amplitude) instead of the start amplitude
and stop amplitude is also possible.
p.90
When the sweep mode is single-shot or gated single-shot, [Trig] (trigger condition) must be set.
p.94
Select [Amptd] in
[Type] and then
press the ENTER
key.
4.8 Sweep Setting and Manipulation