123
IM FG410-01EN
c) Triggeredgateoscillationsettingscreen
1stpage:Screenforthebasicparameters
These are the common setting items that are common to all the oscillation modes.
Press the NEXT
key to display the
2nd page.
Indicates that the 1st
page is displayed.
Set the oscillation
mode to [Burst].
Oscillation start/stop
phase setting.
Oscillation start/stop phase [Phase]
This is the phase at which oscillation starts and stops.
2ndpage:Screenforthetriggeredgateoscillationsettings
Press the NEXT
key to display the
1st page.
Indicates that the 2nd
page is displayed.
Set burst mode to
[
TrigGate].
Stop level
Trigger
Sync output Oscillation stop unit
Sync output
Stop level [StpLvl]
This is the signal level during oscillation stop. The level can be set to either off or on.
Normally, it is set to off.
p.125
Trigger [Trig]
This is the trigger condition.
Internal and external can be chosen as the trigger source.
p.124
Oscillation stop unit [OscStop]
This is the oscillation stop unit. A choice can be made between 1-cycle unit and 0.5-cycle
units. Normally, 1-cycle unit is set.
p.125
Sync output [SyncOut]
This is the output signal from the sync/sub-output terminal. Waveform reference phase
synchronization and burst synchronization can be chosen from.
p.124
d) Waveformsforwhichtriggeredgateoscillationisnotpossible
Gate oscillation is not possible for DC.
Gate oscillation can be done for noise, but the operation differs from that of other waveforms.
p.125
e) Settingitemsrequiredfortriggeredgateoscillation
Set [Phase] (oscillation start/stop phase) on the 1st page of the setting screen.
[StepLvl] (stop level) on the 2nd page of the setting screen is set to [Off] normally.
p.125
[OscStop] (oscillation stop unit) on the 2nd page of the setting screen is set to [Cycle] (1 cycle)
normally.
p.125
A trigger is required for triggered gate oscillation. Refer to the following item.
4.9 Burst Setting and Manipulation