125
IM FG410-01EN
Trigger signal
Gate on
Gate off
Internal gate signal
Burst sync signal
Output signal
Oscillation
Stop Stop
Gate on
Gate off
Oscillation
i) Toexecuteoscillationinhalf-cycleunits:Setoscillationstopunittohalfcycle
To stop oscillation in half-cycle units, set [OscStop] (oscillation stop unit) on the 2nd page of the
setting screen to [HalfCycle] (half cycle). Normally [Cycle] (1 cycle) is set. When [Cycle] (1 cycle) is set,
oscillation is done in integer cycles.,
p.120
j) Usage of stop level
The level when oscillation is stopped is normally set with the oscillation start/stop phase, but it can
also be set independently from that with a proportion of the amplitude full scale.
Set [StpLvl] (stop level) on the 2nd page of the setting screen to [On] (on) and set the level with a %
value.
p.116
k) Noisetriggeredgateoscillation
Noise having no cycle, the gate-on period is the oscillation period, and the gate-o period is the period
stop interval. Further, since noise has no phase, the stop level setting is always enabled.
p.121
4.9 Burst Setting and Manipulation