7. SPECIFICATIONS
The values of items marked with
*1
are guaranteed values. All other values are either nominal values
or typical (typ.) values, and are not guaranteed.
Conditions unless otherwise mentioned are as follows: Continuous oscillation, 50 Ω load, 10 Vp-p/50
Ω amplitude setting, 0 V DC offset setting, auto-range, ±FS waveform amplitude range, external
addition off, AC voltage = RMS value measurement.
7.1 OscillationModes
Continuous, modulated, sweep, burst, sequence
7.2 Waveforms
7.2.1 Standardwaveforms
Types Sine, square, pulse, ramp, parameter-variable waveform, noise
(Gaussian distribution), DC
Polarity Normal, inverted (selectable) (excluding DC)
Amplitude range
-
FS/0, ±FS, 0/+FS (selectable)
(excluding DC)
7.2.2 Arbitrary waveforms
Waveform length 4 K to 512 K words (2
n
, n = 12 to 19) or 2 to 10,000
control points (linear interpolation between control
points)
Total waveform saving capacity Up to 128 waveforms or 4 M words (combined total for
channels 1 and 2)
Saved to non-volatile memory
Waveform data amplitude resolution 16 bits
Sampling rate 120 MS/s
Polarity Normal, inverted (selectable)
Amplitude range
-
FS/0, ±FS, 0/+FS (selectable)
Output bandwidth 25 MHz,
-
3 dB