15
IM FG410-01JA
CAUTION
IfadifferenceinpotentialexistsbetweenthegroundsoftheBNC
connectors,donotshortcircuittheseBNCconnectorgrounds,asthis
maydamagetheequipment.
ATTENTION
Encasdedifférencedepotentielentrelesterresetlesconnecteurs
BNC,nemettezpasencourt-circuitcesterresdeconnecteursBNC:
vousrisqueriezd’endommagerl’appareil.
3.2.1 Waveform output (FCTNOUT)
FG420FG410
CH1
FCTN OUT
[Insulated from housing]
FCTN OUT
[Insulated from housing]
CH2
FCTN OUT
[Insulated from housing]
This is the main output.
Output characteristics
Output voltage Maximum ±10 V/open
Output impedance
50
Ω
Load impedance
0
Ω
or higher (short-circuit protection provided)
Signal GND Insulation from the housing (42 Vpk max.)
FG420: Insulation also between channels (42 Vpk max.)
3.2.2 Sync/sub-output(SYNC/SUBOUT)
FG420FG410
SYNC/SUB OUT
[Insulated from housing]
CH1
SYNC/SUB OUT
[Insulated from housing]
CH2
SYNC/SUB OUT
[Insulated from housing]
A sync signal is output according to the waveform or oscillation status. This signal can be used as the
oscilloscope synchronization signal.
As shown in the following table, the output signal can be selected according to the oscillation mode.
Table 3.1 Signals Selectable for Sync/Sub-Output
Oscillation Mode Selectable Output Signal
All 50% duty TTL level logic signal that rises at zero phase position of reference
phase of waveform output (hereafter, reference phase sync signal).
The phase relationship with waveform output can be changed.
p. 49