MODBUS Communication
15
TI 12A01F01-62EN
l Error settings-sensor
Address
NE107
Address
NE107
Address
NE107
Address
NE107
Name Type
32 F 48 C 64 S 80 M PH Device fail uint16
(bit
elds)
33 49 65 81 Sensor status
34 50 66 82 Measure alarm
35 51 67 83 Device status
36 52 68 84 SC Device fail
37 53 69 85 Sensor status
38 54 70 86 Measure alarm
39 55 71 87 Device status
40 56 72 88 ISC Device fail
41 57 73 89 Sensor status
42 58 74 90 Measure alarm
43 59 75 91 Device status
44 60 76 92 DO Device fail
45 61 77 93 Sensor status
46 62 78 94 Measure alarm
47 63 79 95 Device status
n Error simulation
Simulate the following 4 alarms and output (Error simulate) by the bit.
Measure alarm
I/O alarm
Maintenance status
Setting alarm
ERROR: Provide the alarms with simulated alarm status by bit.
(1: Error on, 0: Error o)
MODE: Dene alarms to simulate by bit.
(1: Error simulate ON, 0: Error simulate OFF)
l Converter error simulation
Address Nane Type
100, 101 MODE Measurement alarm uint32
(bit elds)
102, 103 I/O alarm
104, 105 Mantenance Status
106, 107 Setting alarm
108, 109 ALARM Measurement alarm
110, 111 I/O alarm
112, 113 Mantenance Status
114, 115 Setting alarm
Apr. 19, 2019-00