MODBUS Communication
25
TI 12A01F01-62EN
2.3.8 Display settings Trend (3300-)
Address Parameter Type Description
Trend
1
Trend
2
Trend
3
Trend
4
3300 Process
parameter
int16 pH(1), Temperature(2), ORP(3), rH(4), Conductivity-
TC1(5), Conductivity-TC2(6), Resistivity-TC1(7),
Resistivity-TC2(8), Concentration-TC1(9),
Concentration-TC2(10), Oxygen(11), Di󰀨erential
pH(12), Di󰀨erential ORP(13), Di󰀨erential
CONDUCT(14), Di󰀨erential RESIST(15), Di󰀨erential
DO(16), Average pH(17), Average ORP(18), Average
CONDUCT(19), Average RESIST(20), Average
DO(21), Ratio(22), Passage(%)(23), Rejection(%)
(24), Deviation(%)(25), pH calc.(VGB)(26), AI(27),
Script1(28), Script2(29), Script3(30), Script4(31),
Script5(32), Script6(33), Script7(34), Script8(35)
3306 3312 3318 Empty(0), pH(1), Temperature(2), ORP(3), rH(4),
Conductivity-TC1(5), Conductivity-TC2(6), Resistivity-
TC1(7), Resistivity-TC2(8), Concentration-
TC1(9), Concentration-TC2(10), Oxygen(11),
Di󰀨erential pH(12), Di󰀨erential ORP(13), Di󰀨erential
CONDUCT(14), Di󰀨erential RESIST(15), Di󰀨erential
DO(16), Average pH(17), Average ORP(18), Average
CONDUCT(19), Average RESIST(20), Average
DO(21), Ratio(22), Passage(%)(23), Rejection(%)
(24), Deviation(%)(25), pH calc.(VGB)(26), AI(27),
Script1(28), Script2(29), Script3(30), Script4(31),
Script5(32), Script6(33), Script7(34), Script8(35)
3301 Process
parameter
slot
int16 Sensor1-1(1), Sensor1-2(2), Sensor1-3(3),
Sensor1-4(4), Sensor2-1(5), Sensor2-2(6),
Sensor2-3(7), Sensor2-4(8), Converter(9)
3307 3313 3319 None(0), Sensor1-1(1), Sensor1-2(2), Sensor1-3(3),
Sensor1-4(4), Sensor2-1(5), Sensor2-2(6),
Sensor2-3(7), Sensor2-4(8), Converter(9)
3302,
3303
3308,
3309
3314,
3315
3320,
3321
Low oat arbitrary
3304,
3305
3310,
3311
3316,
3317
3322,
3323
High oat arbitrary
Trend
5
Trend
6
Trend
7
Trend
8
3324 3330 3336 3342 Process
parameter
int16 Empty(0), pH(1), Temperature(2), ORP(3), rH(4),
Conductivity-TC1(5), Conductivity-TC2(6), Resistivity-
TC1(7), Resistivity-TC2(8), Concentration-
TC1(9), Concentration-TC2(10), Oxygen(11),
Di󰀨erential pH(12), Di󰀨erential ORP(13), Di󰀨erential
CONDUCT(14), Di󰀨erential RESIST(15), Di󰀨erential
DO(16), Average pH(17), Average ORP(18), Average
CONDUCT(19), Average RESIST(20), Average
DO(21), Ratio(22), Passage(%)(23), Rejection(%)
(24), Deviation(%)(25), pH calc.(VGB)(26), AI(27),
Script1(28), Script2(29), Script3(30), Script4(31),
Script5(32), Script6(33), Script7(34), Script8(35)
3325 3331 3337 3343 Process
parameter
slot
int16 None(0), Sensor1-1(1), Sensor1-2(2), Sensor1-3(3),
Sensor1-4(4), Sensor2-1(5), Sensor2-2(6),
Sensor2-3(7), Sensor2-4(8), Converter(9)
3326,
3327
3332,
3333
3338,
3339
3344,
3345
Low oat arbitrary
3328,
3329
3334,
3335
3340,
3341
3346,
3347
High oat arbitrary
3348 X-axis:
Timing
int16 15 min.(0), 30 min.(1), 1 hour(2), 2 hours(3), 4
hours(4), 8 hours(5), 24 hours(6), 7 days(7), 14
days(8)