When a training failed
IDnumberindicates
:astatusofUntrained,evenafteranewchromatogramisgenerated,
checkthefollowingitems.
1.Thehealthdiagnosesettingscreenisnotautomaticallyupdated.Youneedtodisplaythe
Listtabagainafterdisplayinganotherscreen.Thedisplayisupdated.
2.HealthdiagnoseisperformedbyusingPCASchromatogram.Therefore,ifthePCAS
foldernameisdierentfromtheoneinthesetting,thechromatogramisnotread.
CheckthePCASfoldernameasfollows.
Select[Conguration]>[SetPCASFolder]onthemenubar.
CheckthefoldernametherematchesthedirectoryoftheactualPCASfolderthatGCAI
refersto.
3.CheckifthefolderwiththecorrespondingIDinthePCASfoldercontainschromatograms
forthepreviousday’sdate.Ifthechromatogramofthepreviousdayisnotstored,the
trainingcannotbeperformedinthedefaultmode.
4.Aftertrainingiscompleted,
remainsdisplayeduntilarstresultcomesout.
CheckwhetheranewchromatogramisgeneratedinthefolderofthecorrespondingIDin
PCASfolder.
5.Iftherearenoproblemsaftercheckingtheaboveitemsandtheautomaticallyregion
settingisconducted,checktheeachregionofthechromatogramleshowsstablevalue.
Ifthevalueisnotstable,itispossiblethatthedefaultregionisnotsetproperly.From
eachID’sheathdiagnosesettingsscreen,gototheregionsettingscreenandcheckto
seeifthereisatleastoneregionthatisenabled.
Ifallaredisabled,youneedtosetuparegionandstarttraining.
6.Otherwise,ifyoudonotidentifythesourceoftheproblem,reviewthelogsinthe
dashboard,andsystemmessagesaswell.
SEE
ALSO
Formoreinformationontheregionsetting,refertothefollowingdocuments:
• “GasChromatographAIMaintenanceSupportHandbook”(TI11B06H01-02EN)Chapter2
• “GasChromatographAIMaintenanceSupportOperatingManual”(IM11B06H01-01EN)Chapter3
TIP
Ifyoudonotusethehealthdiagnosefunction,disableallregions.
TIP
<HowtogeneratePre-anomalyorAnomalyarticially(reference)>
BygeneratinghistoricalchromatogramsinPCASchromatogramfolderfor
(Installation folder) \ PCAS \ data \ (applicable ID) \ (applicable sensor) \ FullCro \ (any date)
youcansimulatethesituationwhereachromatogramisgenerated.Tochecktheobjective,copythedata
withdierencesorpreviouschromatogramsofdatesfarenoughfromthetrainingdates(=theprevious
day)againstthetrainingdatainthedefaultmode,andduplicatetheminthesamefolder.
DuplicatingchromatogramshavenoeectonPCAS