10 - 5
IM 4H3B1-01E
0. GENERAL SPECIFICATIONS
Relation between Chart Speed and Intervals of Periodic Printout:
Pen Model Dot Model
DISPLAY
Display Method: VFD (5 × 7 dot matrix, 20 characters).
Digital Display: AUTO Channel No. or tag, type of alarm, measured values, unit (front 3
digits), for each channel alternately
MAN Channel No. or tag, type of alarm, measured values, unit (front 3
digits), for one specific channel
CLOCK Year / month / date and hour / min / sec will be displayed
VIEW Status for IC Memory Card and TLOG function will be displayed
Bargraph Display: Measured data Left-centered (%) bargraph and center-zero bargraph
(selectable for each channel).
Alarm Setting level of alarm, in case of alarm: flashing display.
Other Displays: RCD: recording in progress
POC: Pen Offset Compensation (pen model only)
SET: SET Mode
AUTO: Auto display is used
MAN: Manual display is used
CLK: Clock display is used
ALM: shared alarm (not corresponding to any channel)
CHT: Chart end indicator (optional)
BAT: low battery
ALARM
Number of Levels: Up to four levels for every channel (High, Low, Delta high, Delta low, High rate-of-
change, Low rate-of-change selectable)
Interval time of rate-of-change alarm is the measuring interval multiplied by 1 to 15
(selectable, and common for both rate-of-change alarms).
Display: Set value: Is indicated as a segment on the bargraph.
In case of alarm: This segment will start flashing on the bargraph and the digital
display will show the type of alarm. The alarm indicator in
different display will show alarm.
Hysteresis: Approx. 0.5% of recording span (only High, Low) and 0%, selectable (common for all
channels and all levels).
Chart Speed Printing Interval of Periodic Printout
6 dot 12 dot 18 dot 24 dot
1 to 9 mm/h no printout no printout no printout no printout
10 to 19 mm/h every 8 hrs. every 12 hrs. every 12 hrs. every 24 hrs.
20 to 39 mm/h every 4 hrs. every 8 hrs. every 8 hrs. every 12 hrs.
40 to 50 mm/h every 2 hrs. every 4 hrs. every 4 hrs. every 8 hrs.
51 to 79 mm/h every 2 hrs. every 4 hrs. no printout no printout
80 to 100 mm/h every hour every 2 hrs. no printout no printout
101 to 1500 mm/h no printout no printout no printout no printout
Chart Speed
Printing Interval of Periodic Printout
5 to 9 mm/h
10 to 18 mm/h
20 to 36 mm/h
40 to 72 mm/h
75 to 135 mm/h
150 to 180 mm/h
200 to 320 mm/h
360 to 1500 mm/h
more than 1500 mm/h
no printout
every 8 hours
every 4 hours
every 2 hours
every hour
every 30 minutes
every 20 minutes
every 10 minutes
no printout