2.2 Appropriate Locations for Using the Instrument
WARNING
Install the instrument so that you can immediately remove the power cord if an abnormal or
dangerous condition occurs.
Do not use the instrument to measure locations that fall under Measurement Categories II,
III, and IV.
CAUTION
This instrument is equipped with voltage and current source and measurement features. Do
not use the instrument when it is wet. Doing so may damage the instrument.
French
AVERTISSEMENT
Installer l’instrument de manière à pourvoir immédiatement le débrancher du secteur en
cas de fonctionnement anormal ou dangereux.
Ne pas utiliser l’instrument pour mesurer des emplacements appartenant aux catégories
de mesure II, III, et IV.
ATTENTION
Cet instrument est doté de fonctions de mesure et de source de courant et de tension. Ne pas
utiliser l’équipement lorsqu’il est mouillé. Le cas échéant, un endommagement de l’équipement
risquerait de se produire.
Installation Conditions
Install the instrument in a place that meets the following conditions.
Flat, Even Surface
Install the instrument on a stable surface that is level in all directions. Pressure may not be measured
correctly when the instrument is placed in an unstable or inclined place.
Ambient Temperature, Ambient Humidity, and Altitude at the
Location of Use
Use the instrument in the following environment.
Ambient temperature 5°C to 40°C (10°C to 35°C when -D00 is selected)
Ambient humidity 20% RH to 80% RH; no condensation
Operating altitude 2000 m or less

Table of Contents