Communication Execution Errors (200 to 299)
Code Message Corrective Action Page
203 Command protected See “Adjustment and Calibration” in section 5.8. 5-12
221 Setting conflict. Check settings that are related to each other. Chapter 11
*
222 Data out of range. Check the ranges of the settings. Chapter 11
*
223 Too much data. Check data byte lengths. Chapter 11
*
224 Illegal parameter value. Check the ranges of the settings. Chapter 11
*
225 OverFlow. Keep program messages to 1024 bytes or less in length,
including <PMT>.
10-2
*
226 Out Of Memory. Keep program messages to 1024 bytes or less in length,
including <PMT>.
10-2
*
241 Hardware missing. Check that the specified options are all installed. —
260 Expression error. Mathematical operations cannot be used. —
270 Macro error. The instrument does not support the IEEE 488.2 macro
specifications.
—
272 Macro execution error. The instrument does not support the IEEE 488.2 macro
specifications.
—
273 Illegal macro label. The instrument does not support the IEEE 488.2 macro
specifications.
—
275 Macro definition too long. The instrument does not support the IEEE 488.2 macro
specifications.
—
276 Macro recursion error. The instrument does not support the IEEE 488.2 macro
specifications.
—
277 Macro redefinition not allowed. The instrument does not support the IEEE 488.2 macro
specifications.
—
278 Macro header not found. The instrument does not support the IEEE 488.2 macro
specifications.
—
* See the user’s manual (IM MT300-01EN), which is included in the accompanying CD.
System Communication Errors (300 and 399)
Code Message Corrective Action Page
300 Communication device-specific error. Servicing is required. —
399 Fatal error in the communication driver. Servicing is required. —
5.3 Communication Error Messages